TRANSCUTANEOUS BILIRUBIN METER

Contract

Model: BILITEST 2000

Producer: TECHNOMEDICA 

Origin: Russia

5875 Viewed

FEATURES:

  • BILITEST - is a portable fully automatic measurement device, intended for objective evaluation of hyperbilirubinemia level in newborns using non-invasive transcutaneous method.
  • Transcutaneous bilirubinometry is a screening method for revealing a group with risks of heavy hyperbilirubinemia progress. Members of this group of risks are to be exposed to further direct measurement of bilirubin, typically with the help of reagentless microanalyser for total blood bilirubin BILIMET C. Use of other laboratory techniques is also available.
  • Use of BILTEST allows reducing the number of newborns subjected to blood collection for laboratory investigations. BILITEST permits thorough control and monitoring in newborns for the jaundice progress and for the efficiency of a treatment course.
  • Commonly in newborns bilirubin determination is performed on a forehead near nose or on the upper part of the breast. Only a slight touch of a mobile measuring head to a babya s derma is needed for measuring. The result is given in international units for transcutaneous bilirubin index (TBI). This index has a high correlation degree with plasma bilirubin concentration (0.92).
  • Power is supplied by 3 AA batteries, the lifetime of a single battery set exceeds 100.000 measuring cycles.

SPECIFICATIONS:

  • Light source: two white light emitting diodes (LED)
  • Method of measurement: photometric reflection.
  • Unit of measurement: TBI (Percutaneous bilirubin index).
  • Corresponding units: mg / dl or umol / l
  • Measuring range: 0-30 mg / dl or 0-500 umol / l
  • Number of spectral channels: 2
  • Correlation coefficient with TSB: 0.92
  • Optical error: <10%
  • Error: ± 2 TBI units
  • Measurement time: 2 seconds
  • Dimensions: 135 x 65 x 35 mm
  • Weight: <150g
  • Power source: 3 AAA batteries
  • Battery life:> 100.000 measurements