Contract
Model: TOP2300
Producer : TOP- Japan
Factory : Meditop - Malaysia
Origin: Malaysia
FEATURES:
- Current speed range.
+ mL / hr (1mL / h increments).
1-999mL / h: TOP-H.
1-300mL / h: 20 drops / mL Standard IV set.
1-150mL / h: 60 drops / mL Standard IV set.
+ drops / minute (1mL / hour increments).
1-333 drops / minute: TOP-H.
1-100 drops / minute: 20 drops / mL Standard IV set.
1-150 drops / minute: 60 drops / mL Standard IV set.
+ Mode microphone (0.1mL / h increments): 0.1 - 99.9 mL / hr (TOP-H series IV set)
- Speed of chasing air: Maximum speed of the set.
- KVO speed range.
+ Normal mode: 1-10 mL / hr (1mL / h increments).
+ Mode micro: 0.1 mL / hour (when set below 0.9mL / hour).
- Limited volume range.
+ Normal mode: 1-9999 mL (1 mL increments)
+ Microphone mode: 0.1 -999.9 mL (0.1 mL increments)
- Total volume range.
+ Normal mode: 0-9999 mL (1mL increments)
+ Microphone mode: 0.0 - 999.9 mL (0.1 mL increments)
- Accuracy of infusion mL / hour.
+ Normal mode: ± 5%: Select set of TOP-H series IV; ± 10%: Select a standard set of 20 drops / mL, 60 drops / mL.
+ Microphone mode: Select the set of TOP-H series IV: ± 5% (custom 1.0-99.9mL / hour); ± 10% (0.1-0.9 mL / h custom).
- Pressure detecting blockage.
+ 4 options (Select the IV series TOP-H).
P13: Under 150kPa Under 1100mmHg Under 1.5 kgf / cm2.
P12: Under 120kPa Under 900mmHg Under 1.2 kgf / cm2.
P11: Under 80kPa Under 600mmHg Under 0.8 kgf / cm2.
P12: Under 40kPa Under 300mmHg Under 0.4 kgf / cm2.
(When the flow rate is 0.9mL / h or less, the device operates P10).
+ 2 options (Select standard IV set of 20 drops / mL, 60 drops / mL).
P21: Under 150 kPa Under 1100mmHg Under 1.5 kgf / cm2.
P20: Under 80 kPa Under 600mmHg Under 0.8 kgf / cm2.
- In-line gas detection: 3 options.
Ar0: High Approximately 50µL.
Ar1: Average Approx. 0.1 mL / 15 minutes.
Ar2: Low Approximately 1mL / 15 minutes.
- When the flow rate is 10mL / h or less, the device operates Ar1 despite setting Ar2.
- Alarm: Internal air, Empty, Speed error, Free flow, Fault wire IV, Clogging, Low battery, Auto power off, Battery connection, Battery problem, Door open, Dropping sensor, Reminder, Check setup, Complete infusion, Switch on power.
- AC power supply: AC 100-240V ± 10%; 50 / 60Hz.
- External DC power: DC 15V ± 10%.
- Internal battery: Rechargeable, DC 12V 1800 mAh Ni-MH (Model BP-23).
- Duration: Approximately 4 hours (At a rate of 25mL / h with a fully charged new battery).